@samuraimittal/node-red-alicat 0.1.1

Node-RED nodes for Alicat mass flow and pressure controllers over Modbus TCP: one supervised socket per device, priority queue, decoded engineering units.

npm install @samuraimittal/node-red-alicat

@samuraimittal/node-red-alicat

CI npm License

Node-RED nodes for Alicat Scientific mass flow and pressure controllers over Modbus TCP:

  • one supervised connection per instrument, shared by every node and kept across redeploys;
  • a priority queue that puts valve commands ahead of polling;
  • readings decoded into engineering units, with a quality flag and a timestamp on every message;
  • setpoints in mass-flow, pressure and valve-drive modes, and a node that maintains a desired state.

The package has zero runtime dependencies.

Status: early release (0.x). Settings and messages may still change between minor versions; the CHANGELOG lists every change.

  • Hardware-tested on a Gen-2 unit (firmware 10v20) and a Gen-1 unit (9v07). Other firmware is covered by the simulator. See docs/hardware-validation.md.
  • Modbus TCP only for now; Modbus RTU and serial connections are planned.

Not affiliated with or endorsed by Alicat Scientific, Inc.

Contents

Install

Requires Node-RED 4.0 or newer and Node.js 18.5 or newer, as Node-RED 4 does (Node-RED 5 itself needs Node.js 22.9+).

In the Node-RED editor: Menu → Manage palette → Install, search for @samuraimittal/node-red-alicat. Or from a shell, in your Node-RED user directory (usually ~/.node-red):

cd ~/.node-red
npm install @samuraimittal/node-red-alicat

Restart Node-RED. The nodes appear in the palette under Alicat.

Quick start

Without hardware: the package ships a simulator that behaves like a real 10v20 controller.

cd ~/.node-red
npx --no -- alicat-sim --port 5020     # a second one for example 05: npx --no -- alicat-sim --port 5021
npx --no -- alicat-sim --list          # other profiles, e.g. --profile 9v07-bench for Gen-1 firmware

--no makes npx use the installed simulator and never download a package of that name.

In Node-RED, use Menu → Import → Examples → @samuraimittal/node-red-alicat → 01-poll-and-dashboard, then deploy. The examples point at 127.0.0.1:5020 (and 5021 for the two-station example).

With an instrument:

  1. Drag an alicat read node onto the flow and open it.
  2. Next to Device, click the pencil to add a device. Enter the instrument's IP address and keep port 502.
  3. Click Test connection. It shows the firmware, what the setpoint controls, and live readings.
  4. Deploy. The node's status shows the live mass flow, for example 12.47 SLPM.

For firmware before 10v07 (Gen-1), also pick the data frame layout in the device settings. extended-7 matches the common 7-slot frame (absolute, gauge and barometric pressure, temperature, volumetric flow, mass flow, setpoint).

The nodes

Node What it does
alicat device (config) One instrument: address, port, unit ID, and for Gen-1 firmware the data-frame layout, controlled variable and unit labels. Advanced: timeouts, request gap, heartbeat, retry limit, frame trace.
alicat read Live readings on an interval or on input, as one object or a single number. Optionally publishes a frozen snapshot to flow/global context.
alicat setpoint Writes a setpoint and confirms it by reading it back. Can switch what the setpoint controls (mass flow, volumetric flow, absolute/gauge/differential pressure, valve drive) without a jump.
alicat command Tare, totalizer reset, valve hold closed / hold current / exhaust / resume, gas select, front-panel lock. Each command is confirmed by the device.
alicat PID Reads or writes the control-loop gains, or sets the loop algorithm.
alicat status Connection and device state for alarms and dashboards; an input can force a reconnect.
alicat control Keeps the instrument at a desired setpoint, mode and valve state, and corrects drift. Never acts on stale data.

Every node has full help in the editor's help panel. The examples cover each node:

Example Shows
01-poll-and-dashboard Polling, single readings, identity, status
02-setpoint-control Setpoints, mode switching, an out-of-range value going to Catch
03-tare-totalizer-gas Tare, totalizer reset, gas by name or number
04-pid-read-write PID read, write with read-back, loop algorithm
05-multi-station-floor Two instruments in one flow, one connection each
06-connection-alarms-and-safe-state Link alarms and an end-of-test safe-state sequence
07-migrate-from-globals Replacing a global-variable driver: context snapshot, link call, maintained setpoint

Compatibility

Firmware Support
10v07 and newer (Gen-2) Identified automatically: readings, units, ranges and control mode come from the device. Extended 32-bit commands.
10v19 and newer Also serial number, dates, the control-mode query (command 43) and totalizer 2 reset.
6v17 to 10v06 (Gen-1) Readings from the device's data frame. The layout, controlled variable and unit labels are set in the device configuration, because this firmware can't report them. Limited command set.
Hardware-tested 10v20.0 and 9v07, on native Ethernet.
Not supported BASIS 2 and CODA instruments (different register maps); serial gateways and Modbus RTU.
Runtime Verified with
Node-RED 5.0.8, Node.js 24 (Windows 11) Full test suite, bench hardware, 24-hour soak (results)
Node-RED 4.1.16, Node.js 24 Full test suite
Node-RED 4.1.16, Node.js 18.20 Runtime smoke test (the test tools themselves need Node.js 20+)
Node.js 22, Linux In the CI workflow (Node-RED 4.1 and 5.0 on Node.js 22 and 24, Linux and Windows)

Status dots

Dot Text (examples) Meaning
yellow ring connecting Opening the connection
yellow dot identifying Connected; reading firmware and layout
green dot 12.47 SLPM, SP 12.5 SCCM, tare ✓ Ready; the node's latest value or result
yellow dot tare: rejected The last operation failed, but the link is fine
red ring retry 3: refused Not connected; retrying with backoff
red dot duplicate device Another configuration already uses this instrument
red dot set layout, set device type, check unit ID, not an Alicat? Connected, but identification needs a setting
grey ring closed Stopped

Messages

Every output carries msg.alicat:

Property Meaning
device, host, port, unitId Which instrument
generation, firmware, serial gen1 / gen2, e.g. 10v20.0, and the serial number (10v19+)
op Which node sent it: read, setpoint, command, pid, status or control
quality good, stale (the device stopped answering; values are the last good ones) or offline
ts, ageMs When the reply arrived (ms since epoch), and how old the values are
units Unit of each reading, e.g. {massFlow: "SCCM", gaugePressure: "PSIG"}; the setpoint node sends {setpoint: "SCCM"}
status Decoded status bits: {raw, active, flags}; e.g. flags.holding, flags.exhausting
gas, controlledVariable, setpoint Current gas, what the setpoint controls, the setpoint
outcome Setpoint, command and PID nodes: the operation's outcome (see below)
result The operation's details: the setpoint write and read-back, the command's device status, the status node's identity
error On failure: {code, message, outcome, …}, also seen by Catch nodes. It replaces msg.alicat, so nothing from an upstream node is mixed in

A node that has no device selected, or invalid settings, shows a red status and fails every message with EINVALIDARG (outcome not-sent), so the problem reaches Catch nodes instead of messages vanishing.

When an instrument stops answering, a read node sends one message with quality: "stale" or "offline" and then waits. It never keeps sending frozen values as if they were live. Check msg.alicat.quality === "good" before using a value for a control decision.

Outcomes:

Outcome Meaning
executed Confirmed by the device (echo, read-back, or the effect on the status bits)
accepted The device took it, but the result can't be read back (e.g. a mode switch on Gen-1)
rejected The device refused it (ECOMMAND, with the device's status)
unknown It may have been sent, but nothing proves the result (timeout after the write, pre-emption, another master)
not-sent Nothing was written

Errors

Failures go to Catch nodes. msg.alicat.error.code is one of a fixed set (it is part of the public API):

Code Meaning Usual cause
ECONNECT Could not connect Wrong IP, device off, cable, firewall
ENOTCONNECTED Not connected The link is down; writes fail fast instead of running later
ELINKDOWN Connection lost Cable pulled, device restarted
ETIMEDOUT No response Device busy or unreachable
EMODBUS Modbus exception from the device Register not available on this firmware; device busy
ECOMMAND The device rejected a command Unsupported argument (e.g. differential pressure without that sensor)
EPREEMPTED An urgent valve command interrupted this command Expected when a hold is sent during a tare
EINVALIDARG Invalid input Setpoint not a number, out of range, unknown gas
EUNSUPPORTED Not available on this firmware e.g. command 43 before 10v19
EIDENTIFY Could not identify the device Gen-1 layout not set, wrong device, unit ID
EDUPLICATE Another configuration already uses this device Two device configurations for one IP
EQUEUETIMEOUT, EQUEUEFULL The request could not be sent in time / too many waiting Too much polling for the link
ECANCELED, ECLOSED Cancelled because the node or flow stopped Redeploy or stop during a request
EFRAMING Malformed Modbus frame Not an Alicat, or a gateway in the path

The operating-system error (for example ECONNREFUSED) is in the log and in err.cause.

Deploys, stop and safety

  • One connection per instrument. All nodes using a device share it. Redeploying, even repeatedly, keeps the same connection: the instrument never sees a reconnect. A second device configuration for the same IP is refused instead of opening a second socket.
  • Stop flows and shutdown close the connection immediately, so the instrument's connection slot is free for other masters.
  • The instrument is left as it is on deploy, stop and shutdown: setpoint, mode and valve state don't change. Node-RED is not a safety controller, and Alicat's communications watchdog is not available over Modbus TCP. Put the equipment in a safe state with your own sequence (see example 06) and with independent safety hardware.
  • alicat control starts empty after a deploy or restart: it does nothing until it receives a desired state.
  • Hold closed with fallback (opt-in on alicat command): if the instrument is still finishing another command, setpoint 0 is written at once as well.
  • One master. The command registers are shared between all masters. If a PLC or HMI also sends commands, results can be reported as unknown, and the log warns about another master.

Modbus TCP has no authentication. See SECURITY.md for network and editor guidance.

Troubleshooting

Symptom What to check
retry N: timeout IP address and subnet. Instruments with a link-local address (169.254.x.x) need the PC on the same link. The first connection after plugging in can time out once while Windows sets up the adapter; it retries by itself.
retry N: refused Port 502, and whether the instrument's Modbus TCP is enabled. If it worked before, another master may be holding a connection slot.
duplicate device Two device configurations point at the same instrument (perhaps by IP and by host name). Use one.
set layout Gen-1 firmware: pick the data-frame layout matching the instrument's front-panel data frame.
not an Alicat? Wrong IP or port, or a device that isn't an Alicat on native Ethernet.
Readings stale now and then Too much traffic for the link (many fast read nodes), or network drops. One read loop serves all read nodes of a device, at the fastest interval any of them asks for.
A setpoint seems ignored The setpoint's meaning depends on the control mode (msg.alicat.controlledVariable): 50 in valve-drive mode is 50 %, not 50 SCCM.
"not a private address" in Test connection The IP is outside the private ranges (e.g. 172.168.x.x is public; private is 172.16–31.x.x). Fix the IP plan before routing that network.

The Test connection button in the device configuration identifies the instrument before you deploy, and reuses the live connection if there is one. It only opens new connections to port 502; to test an instrument or simulator on another port, list the port in settings.js as alicatModbus: {probePorts: [5020]} (npm run dev does this for its simulators). Frame trace (Advanced) logs every Modbus frame.

Moving away from global variables

Flows that poll a device into many global variables, and drive it by writing a target into a global, can move over gradually (example 07):

  1. alicat read publishes one frozen snapshot per instrument to context ({ts, quality, values, status, units, …}), from a single writer. Existing readers keep working; they should check quality and age.
  2. Send setpoints and valve states to alicat control through link call, so the caller gets a result or an error instead of polling a global.
  3. Feed analytics (smoothing, slopes, peaks) from messages, using msg.alicat.ts for time.

Developing

npm install
npm run check          # lint, typechecks, tests with coverage gate, runtime smoke test, package contents
npm run dev            # Node-RED on http://127.0.0.1:1880 with this package linked, plus simulators on 5020-5022
npm run sim            # simulated 10v20 MFC on 127.0.0.1:5020 (drops idle links after 20 s, like the real unit)
npm run sim -- --list  # available simulator profiles
npm run probe -- 192.0.2.10 info   # bench probe for a real unit (writes need --write)
npm run test:compat    # runtime smoke test, needs only Node-RED (CI: Node 18 with Node-RED 4.0 and 4.1, and the packed tarball)
npm run soak -- --hours 24 --real 192.0.2.10   # long-running soak test, see test/soak/soak.js
npm run diagram        # rebuild the architecture diagram in docs/diagrams/ after changing tools/diagram/spec.js

With npm run dev, import the examples from Menu → Import → Examples → @samuraimittal/node-red-alicat. Examples use the simulators on 5020 and 5021; 5022 is a Gen-1 (9v07) replica that needs the extended-7 layout. The Node-RED user directory is .dev/node-red; delete it to start fresh.

Conventions, tests and the hardware rules for contributors: CONTRIBUTING.md. Hardware checks (read-only unless enabled) are described in docs/hardware-validation.md. How the pieces fit together: docs/architecture.md, with a diagram of every file and the main flows (docs/diagrams/, editable in Excalidraw). The register map as verified on hardware: docs/register-map.md.

Layout

Path Contents
nodes/ The seven Node-RED nodes (runtime .js plus the editor template .html); nodes/locales/en-US/ holds each node's messages (.json) and help (.html)
resources/ Editor scripts (one per node, plus shared validators) and styles, served to the Node-RED editor
examples/ Importable flows, one per use case; tested against the simulator
lib/node-red/ Node glue: defineDeviceNode (settings, input pipeline, close order), the settings bounds table, the msg.alicat contract, status dots, editor endpoints (/alicat-modbus/probe, gases, layouts)
lib/Reconciler.js The alicat-control engine: holds a desired setpoint, mode and valve state, corrects drift, never acts on stale data
lib/ContextPublisher.js Optional single-writer, frozen snapshot in flow/global context, for flows that still read globals
lib/AlicatDevice.js One instrument: state, identification lifecycle and the public API; the work is in lib/device/
lib/device/ Identification (Gen-1 or Gen-2), snapshot decoding, the shared poll loop, setpoints and mode switches, valve, gas, tare and PID commands
lib/CommandLane.js One command at a time per device, urgent pre-emption; lib/lane/ runs the handshake measured on hardware (NOP first, device-cycle settle, echo polling) for the extended and limited interfaces
lib/AlicatClient.js One supervised connection per device: state machine, single in-flight request, pacing, heartbeat (5 s, inside the device's ~20 s idle drop), link-dead detection, jittered backoff, rate-limited logs
lib/PriorityQueue.js URGENT / CONTROL / POLL queue: coalescing, latest-wins, per-waiter deadlines, POLL aging, bounded depth
lib/ConnectionRegistry.js Process-wide owner of connections: one socket per device, parked across deploys, alias and duplicate detection
lib/modbus/ Modbus TCP transport over net: MBAP framing, 16-bit transaction IDs, late-reply handling
lib/alicat/ Register map constants (1-based, as in the Alicat manual), codec, firmware gates, gas, unit and status tables
lib/errors.js Closed set of err.code values (public API)
lib/util/ Small shared helpers: own-key lookups, defaults, timers
tools/ alicat-sim (simulator, shipped), alicat-probe (bench probe, shipped), dev (Node-RED + simulators), check-pack (package contents), architecture-diagram with diagram/ (builds docs/diagrams/)
tools/sim/ Alicat simulator: process model, register bank, Modbus TCP server, fault injection, test hooks
test/ Unit, integration and node tests (mocha + node-red-node-test-helper) with shared helpers/, compat/ runtime smoke test, soak/ long-running test, hil/ opt-in hardware tests

License

Apache-2.0. Alicat is a trademark of Alicat Scientific, Inc.; this project is not affiliated with or endorsed by Alicat Scientific.